Metal conductivity measurement device PCE-COM 20
$7,755.00
The metal conductivity meter, PCE-COM 20, uses eddy current to determine the conductivity of metals such as aluminum, copper, titanium, magnesium or bronze.
– Operating frequency: 60 kHz
– Sine wave probe: Ø 14 mm / ≈ 0.55 in
– Memory: Up to 500 groups of measurements
– Lift-off and temperature compensation
– Automatic calibration
Metal Conductivity Measurement Device, PCE-COM 20
The metal conductivity meter, PCE-COM 20 uses eddy-current to determine the conductivity of metals such as aluminum, copper, titanium, magnesium or bronze.
HIGHLIGHTS
- Eddy current measuring principle
- For non-ferrous metals and alloys
- Non-destructive testing (NDT)
- 60 kHz operating frequency
- Large measuring range up to 112% IACS or 65 MS/m
- Memory for up to 500 groups of measurements
- Automatic temperature compensation (ATC)
- Lift-off compensation up to 500 μm
- Automatic calibration
- Adjustable backlighting
- Internal rechargeable battery
- Portable, handheld
- User friendly
- Samples of Titanium 1.03% IACS, Bronze 8.11% IACS, and Copper 100% IACS
- For non-ferrous metals and alloys
- Non-destructive testing (NDT)
- 60 kHz operating frequency
- Large measuring range up to 112% IACS or 65 MS/m
- Memory for up to 500 groups of measurements
- Automatic temperature compensation (ATC)
- Lift-off compensation up to 500 μm
- Automatic calibration
- Adjustable backlighting
- Internal rechargeable battery
- Portable, handheld
- User friendly
- Samples of Titanium 1.03% IACS, Bronze 8.11% IACS, and Copper 100% IACS
SPECIFICATION
Operating frequency | 60 kHz, sine wave |
Conductivity measuring range | 0.51% IACS ... 112% IACS 0.3 MS/m ... 65 MS/m |
Resistance range | 0.015388 ... 3.33333 Ω • mm² / m |
Conductivity measuring resolution | 0.01% IACS (at <51% IACS) 0.1% IACS (at 51% IACS ... 112% IACS) |
Conductivity measuring accuracy | ± 0.5% at 20°C / 68°F ± 1% in the range 0 ... 40°C / 32 ... 104°F |
Lift-off compensation | 0.5 mm / ≈ 0.0197 in |
Temperature measuring range | 0 ... 50°C / 32 ... 122°F |
Temperature measuring accuracy | ± 0.5° |
Temperature compensation | Automatic adjustment of conductivity result to the value at 20°C / 68°F |
Operating conditions | 0 ... 50°C / 32 ... 122°F, 0 ... 95% relative humidity (RH) |
Display | LCD, backlit |
Menu languages | English, German, Chinese (simplified) |
Power supply | Internal, rechargeable battery |
Probe diameter | 14 mm / ≈ 0.55 in |
Memory | Stores up to 500 groups of measurement values |
Data interface | USB port connection |
Dimensions | ≈ 220 x 95 x 35 mm / 8.66 x 3.74 x 1.38 in |
Weight | ≈ 415 g / 1 lb (with probe) |
DELIVERY SCOPE
1 x PCE-COM 20 conductivity tester for metals
1 x Sine wave measuring probe
1 x Set of calibration plates (Titanium 1.03% IACS, Bronze 8.11% IACS, and Copper 100% IACS)
1 x PC-compatible software
1 x USB to RS-232 data cable
1 x Charging adapter
1 x Hex wrench
1 x Carrying case
1 x User manual
1 x Sine wave measuring probe
1 x Set of calibration plates (Titanium 1.03% IACS, Bronze 8.11% IACS, and Copper 100% IACS)
1 x PC-compatible software
1 x USB to RS-232 data cable
1 x Charging adapter
1 x Hex wrench
1 x Carrying case
1 x User manual